HOME

 Experimental apparatus

 Some experimental apparatus available in the ESR group are introduced.

☆Low frequency ESR(For absorption spectra)

The low frequency ESR apparatus is useful for measuring spin susceptivility of organic substances, especially with Pauli paramagnetism. As you know, SQUID susceptometer is superier in sensitivity, but it cannot descriminate Pauli paramagnetism from the diamagnetic contributions of the substance, both of which are usually temperature independent. However, ESR observes only spin magnetisms, but not diamagnetisms, which is a clear advantage over the SQUID susceptometer.
One usually uses an electromagnet with iron cores: ESR is measured at low magnetic field and then, its intensity is calibrated with 1H NMR signal intensity measured simply by increasing the magnetic field, which satisfies the NMR resonance condition. We call these procedures as the ESR-NMR technique. This is a good way, but it is a relatively time consuming procedure, There are mainly two reasons:

  • To keep accurate, the amplitude of the field modulation to observe a derivative signal of the absorption spectrum should be sufficiently lower than the spectral line width, resulting in the suppression of sensitivity in ESR and NMR measurements. As a result, lots of accumulations of signals are required for us to achieve an enough S/N ratio.

  • There is a significant limitation in the measurement conditions: Magnetic field scans to accumulate signals should be enough slow to keep the scan rate unchanged because of the iron hysteresis of the electromagnet with iron cores.

The low frequency ESR system shown in the above photographs has several advantages over the usual ESR-NMR technique. Characteristics are summarized as follows.

  1. Measures only ESR as functions of temperature and pressure, and calibrate by comparing it with that of a standard sample, for example, DPPH.

  2. Magnetic field is produced by an empty electromagnetic coil without any cores, which is basically independent of the hysteretic distortions in ESR baselines, even in fairly rapid scan rates of the magnetic field.

  3. ESR absorption spectra are measured without the magnetic field modulation.

The rapid scan rates make us possible to measure the absorption spectra without the field modulation, which is a requirement to suppress the 1/f noise. We can use the scan rate higher than 10 Hz without any distortion of the signals. Since both of the sample and the standard sample are measured in a sample tube simultaneously, the linewidth of the standard sample should be sufficiently differrent from that of the sample. If the number of spins in the samples are the same as each other, the intensity ratio of the ESR absorption signals is proportional to the ratio of the linewidths, which is markedly different from the square of the linewidth ratio in the derivative signals with the magnetic field modulation.

As a result of the above factors, the consumption time of the measurements with the present system is much less than that with the ESR-NMR technique, 1/10-1/100 or more.

  • Intensity of the absorption in the sharp signal is more than three times larger than that of the derivative signal with the modulation, since the amplitude of the modulation should be less than 1/3 of the linewidth. Intensity of the broad signal measured with the same modulation amplitude as the sharp one is further weak by the ratio of the linewidth, which corresponds to more than 1/10 in total.

  • Scan duration for the ESR-NMR technique is around 60 s for up scan, which is 600×2 times longer than the present system with a 10 Hz scan for up and down scans. Thus, the final ratio of the S/N ratios reaches square root of 1,200, that is ~1/35.

Thus, this system saves the measurement time by 1/10×1/35. The magnetic field coil is located in a liquid nitrogen space to reduce the Joule heating, which is driven by a 60V-2.5A constant current power supply in the lowest part of the rightmost picutre. ESR frequency is 150-200 MHz with the magnetic field of 5-7 mT.
Development of a similar system for the cubic amvil cell up to 8 GPa is under way.


☆STM & AFM

STM or AFM microscope made by Nanosurf™, Switzerland applicable under ambient conditions. The above-left picture is set for STM microscope. By disconnecting the head part with the circular cover from the long and large controller part, an AFM head can be connected instead, on an active vibration isolator. A computer with a measurement application is connected to an USB port of the controller. Since STM head is constructed so rigidly, STM image is insensitive to ambient vibrations without the vibration isolator.

Atomic scale images of HOPG (Highly Oriented Pyrolitic Graphite) surface can be easily and reproducibly obtained in ambient circumstances with a selfcut probe tip made of Pt-Ir wire. The above-right image was taken by STM in the 2.8×2.6 nm2 range of the a-b plane of α-(BEDT-TTF)2I3 single cristal. Each white island corresponds to the Sulfer π-orbital(s) in a BEDT-TTF molecule. End ethylene group of the molecule could not be observed by STM, because the ethylen contains no electronic states near the Fermi surface of the crystal. Four inequivalent molecules are confirmed in a unit cell of the crystal with this high resolution image.


☆Glove Box

 Since we treat mainly organic materials, such as conducting polymers doped with halogens and alkaline ions, and fullerens doped with strong acids and alkaline ions, glove box is an inevitable apparatus to prepare and treat the samples for several measurements like ESR and electrical conductivity under ambient condition or high pressure up to 2 GPa. The glove box with the helium gas keeps the ambient atmosphere free from oxygen and moisture by circulating helium gas through the refining columns for oxygen and moisture. Inside of the two metal covers, two gloves are equipped, with which one treats samples with a precise weight and a video microscope monitored by a liquid crystal display inside the box. The concentration of oxygen and moisture is controlled in the level of ppm (parts per million), where the moisture never condensates down to minus 100 ゚C.